Products

Medical Pump

 

Sales points

1. Dual screen, double function. Equipped with infusion function and syringe function.

 

2. With LED light, can use in various lighting conditions. 

 

3. Data Memory Functionsave the last setting of sound volume level, brand and occlusion

pressure level, to prepare for the next use. 

 

4.Bubble Alarm: when bubble length is about 3mm (10.6ul) or accumulated bubble length is about

3mm (10.6ul) is detected, will trigger acoustic-optic alarm.

 

5. Power Supply Switching Function: When the AC power supply is off, it will automatically switch to internal battery power supply and prompt at the same time. When AC is switched on, it is automatically converted back to AC supply.

 

6. Keep-Vein-Open (KVO): After finishing predetermined infusion volume, to prevent the formation of thrombus, continue infusion with preset minimum rate 0.1-5ml/h(setted).

 

7. Move Alarm: When user tries to remove or take away the syringe, give move alarm.

 

8. Drip Sensor Slope: In d/min mode, when there are problems such as tube slope, liquid

overflow and misplaced/moved sensor, there will be an alarm.

 

 

Basic Parameter

 

Item

Model

Sun-590

Dimensions(mm)

354(L)×132(W)×199(H)

Weight

2.7kg

Battery Continuous Use Time

About 6 hours

(fully charged new battery running with rate 5ml/h)

Rated Consumption

≤25 VA

 

 

 

 

 

 

 

 

 

Power supply

AC 100V~AC 242V,50/60HZ(±1 HZ)

Built-in battery

DC 12V 1800mAh, Ni-MH, rechargeable

Infusion Flow Selection

0.1ml/h~1200ml/h (1-150d/min), original value is 0.1ml/h (1d/min)

Syringe Flow Selection

0.1ml/h~1600ml/h, original value is 0.1ml/h

Syringe Size

5/6ml, 10/12ml, 20/25ml, 30/35ml,50/60ml

Bolus rate

Syringe 50ml:1600ml/h

Syringe 30ml:900 ml/h

Syringe 20ml:600 ml/h

Syringe 10ml:400 ml/h

Syringe 5ml200ml/h

Infusion Range

0.1ml~9999.9ml, original value is 0.1ml

Infusion Precision

volume mode ±5%, drop mode ±2d

Syring Flow Rate Precision

±3%

KVO

0.1~5ml/h

Occlusion Pressure

High:110kPa±20kPa (825mmHg±150 mmHg)

Middle:80kPa±30kPa  (600mmHg±225mmHg)

Low:60kPa±20kPa  (450mmHg±150 mmHg)

The biggest output pressure

975mmHg(130kPa)

Use Environment

Temperature:  5~43

Relative Humidity:90%RH (Under without condensation status)

Atmospheric Pressure:860hPa~1060hPa

Storage Environment

Temperature:-20~+55

Relative Humidity:90%RH (Under without condensation status)

Atmospheric Pressure:860hPa~1060hPa

Waterproof Level

IPX4

Electric Shock Protection

Class I, Type CF

Alarm

occlusion, nearly finish, finish, standby (no operation),

door open, low battery, battery exhausted etc.

Function

power indicator, transfusion amount display, flow rate

display, infused volume display etc.

 

 

 

1. Package: double carton boxes     

 

2. Port: Shenzhen China, Shanghai China 

 

3. Delivery time: 3-5 working days after receipt of payment

 

4. Payment term: T/T in advance

Back Up