Introduction:
This equipment can monitor such parameters as ECG,RESP,SpO2,NIBP,and TEMP. It integrates parameter measuring module,display and recorder in one device to form a compact and portable device. At the same time, its built-in replaceable battery provides convenience for patient moving. Modular designed, easy to maintain and replaced.
Features:
10.4 inch high definition color TFT display, touch screen.
Standard 6-parameter : ECG, HR, NIBP, SPO2, RESP, TEMP.
13 kinds of Arrhythmia analysis.
2-4 hours battery power supply.
720 hours of trend data playback.
Sound, light, text alarm, volume adjustable.
Connected to central monitoring system .
Optional thermal recorder, ETCO2 module, IBP module,Tolley .
Monitoring contents, scan speed, volume and output contents can be set optionally.
Function of NIBP review, storage for up to 2400 NIBP data.
Adopt digital SpO2 technology, which has strong anti-interference and antiweak filling capability.
Calculation of drug concentration.
Network: connecting with central station, other Bed observation and software updating. Connection mode: wireless and wired.
Built-in rechargeable battery for uninterrupted monitoring.
Print ECG, SpO2, RESP, BP and temperature data with one-key.
Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).
Analysis function for heart rate variability(HRV) (optional)
Performance:
ECG
|
Lead Mode 3-lead and 5-lead are optional Lead Selection I, II, III, avR, avL, avF, V Wave 5-lead: 2 channels 3-lead: 1channel Gain ×2.5mm/mV, ×5.0mm/mV, ×10mm/mV, ×20mm/mV HR Measuring and Alarm Range Range 15 ~ 300 bpm Accuracy ±1% or ±1bpm, which is greater Alarm Accuracy ±2bpm Resolution 1 bpm |
CMRR
|
Monitor ≥ 100 dB Surgery ≥ 100 dB Diagnosis ≥ 60 dB |
Bandwidth
|
Surgery 1 ~ 20 Hz(+0.4dB,-3dB) Monitor 0.5 ~ 40 Hz(+0.4dB,-3dB) Diagnosis 0.05~75Hz(+0.4dB,-3dB); 76Hz~150Hz(+0.4dB,-4.5dB) Calibration Signal 1 mV (Vp-p), ±5% Accuracy |
ST Segment Monitoring |
Measuring and Alarm Range -0.6 mV~ + 0.8 mV |
ARR |
ARR Detecting Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC |
Alarm |
Available |
Review |
Available |
Scan Speed for ECG Waveform is adjustable |
12.5mm/s accuracy ±10% 25mm/s accuracy±10% 50mm/s accuracy±10% |
Respiration
|
Method R-F(RA-LL) Impedance Differential Input Impedance >2.5 MΩ Measuring Impedance Range 0.3~5.0Ω Baseline Impedance Range 100Ω– 2500Ω Bandwidth 0.3 ~ 2.5 Hz |
Resp. Rate
|
Measuring and Alarm Range 0~120rpm Resolution 1 rpm Measuring Accuracy ±2 rpm Alarm Accuracy ±3rpm Apnea Alarm 10 ~ 40 S |
NIBP |
Method Oscillometry Mode Manual, Auto, continuous Measuring Interval in AUTO Mode 1 /2 /3 /4 /5/ 10/ 15/ 30 /60 /90/120/240/480/960 Min Measuring Period in Continuous Mode 5 Min Measuring and Alarm Range 10~270mmHg Alarm Type SYS, DIA, MEAN |
Resolution
|
Pressure 1mmHg Cuff Pressure ±3 mmHg Accuracy ±10% or ±8mmHg, which is greater Over-pressure Protection: Adult Mode 315±10 mmHg Pediatric Mode 265±10 mmHg Neonatal Mode 155±10 mmHg |
SpO2 |
Measuring Range 0 ~ 100 % Alarm Range 0 ~ 100 % Resolution 1 % Accuracy 70% ~ 100% ±2% 0% ~ 69% unspecified |
Pulse Rate(PR)
|
Measuring and Alarm Range 0~250bpm Resolution 1bpm Measuring Accuracy ±2bpm or ±2%, which is greater Alarm Accuracy ±2bpm |
TEMP |
Channel dual-channel Measuring and Alarm Range 0 ~ 50°C Resolution 0.1°C Accuracy ±0.1°C Actualization Interval about 1 Sec. Average Time Constant < 10 Sec. Alarm responding Time ≤2min |
EtCO2 |
Method Sidestream or Mainstream Resolution for CO2: 0.1 mm Hg 0 to 69 mm Hg Apnea Alarm Available |
IBP |
Channel dual-channel Label ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measuring and Alarm Range -50~350 mm Hg Accuracy ±2% or 1mm Hg, which is greater |
Dsplay Mode 12.1" color TFT LCD with high-resolution.
Power Supply 220V, 50Hz
Safety Classification class Ⅰ,type CF defibrillation-proof part
Physical Characteristic: Dimension 380×350×300(mm) Net weight 4.8Kg
Accessories:
1) Adult SpO2 probe(5-pin) 2) Adult NIBP cuff 3) Extending tube for blood pressure 4) ECG lead
5) ECG electrode 6) Temperature probe 7) Power cord 8) Thermal recording paper(optional)9) User Manual