Products

Medical Pump

 

 

SUN-900 Plus Infusion Pump

Display

4.3" LED color screen with 272×480 display

Three operating modes

Rate Mode, Time Mode and Volume Mode

New features

NIBP and SpO2

Drug library

with 210 kinds of medicine

Heating function

Suitable for infusing in winter or there is requirement for 

the fluid medication temperature. 

Remote Controller

Remote Controller makes operating more convenient.

Central monitoring system

Central monitoring system is optional, the infusing status 

of each pump is displayed in real time at the central station 

through wireless transmission.

Operation

Simple and convenient in operating, intuitive presence of working 

status.

Alarm

Audible and Visible alarms

 

Alarm: Infusion Complete, Empty, Faulty Signal, 

Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, 

Setting Error, AC power off, Idle.

Storage

1500 history events

Language

Support multi language display

Size

Size: 188 mm(L)*198 mm(W)*228 mm(H)

Weight

Net Weight: 2.2kg

 

 

 

Technical Specifications

 

Adjustable Drop Rate Range

1~400 drops/min (step: 1 drop/min)

Adjustable Flow Rate Range

1~1200 ml/h (step: 0.1 ml/h)

Purge Rate

800 ml/h

KVO Rate

1~3 ml/h 

Infusion Volume Range

1 ml~9999.9 ml (step: 0.1 ml)

Maximum Accumulated Volume

9999.9 ml (step: 0.1 ml)

 

Time Range

1 min~9999 min (step: 1 min)

Transfusion accuracy

Within 3% (to choose the high quality 

and had calibration apparatus infusionis)

Occlusion Alarm 

Lower: 300mmHg ± 100mmHg

Medium: 500mmHg ± 100mmHg

High: 800mmHg ± 200mmHg

Alarm

Occlusion,Air, Door open,Infusion Complete,

Signal error,Battery undervoltage, Net, power off,

 Empty bottle, Low voltage, Error setting

Air Bubble Detector Method

ultrasonic wave, sensitivity:  ≥ 25 μL

Fuse

F1AL/250V ("F" indicate fast,"L" 

indicate low breaking capacity), 

2 pcs (installed inside pump).

Power Supply

AC 85~265V, 50/60Hz;

Internal Battery

11.1V rechargeable li-ion battery. 

Capacity: ≥ 
2000mAh;

the pump can work more than 

4 hours at the flow rate of 

25 ml/h after charging for 8 hours 

(Medium rate specified by GB 9706.27-2005).

Power

≤ 40VA

Size

188 mm(L)*198 mm(W)*228 mm(H)

Weight

2.2KG

Environment Requirements

Transport

Ambient Temperature: -30~+55

Relative Humidity: 20%~95% (non-condensing) 

Atmosphere Pressure: 70kPa~106kPa

Storage

Ambient Temperature: -30~+55

Relative Humidity: 20%~95% 

Atmosphere Pressure: 70kPa~106kPa 

Operation

Ambient Temperature: +5~+40

Relative Humidity: 20%~90% 

Atmosphere Pressure: 86kPa~106kPas

Calibration function

disposable IV set of any brand is 

suitable for this pump

Three working modes

Rate/Volume/Time mode

Alarm

Infusion Complete, Empty, Faulty Signal, 

Misoperation, Occlusion, 

Door Open, Air Bubble, Low Battery, 

Setting Error, AC power off, Idle.

Infusion monitoring system (Optional),

the status of each pump is displayed in real time at the central station 

through wireless transmission.

4.3 inch color LCD display, with backlight, suitable for working in various ambient light conditions.

Display time, battery, infusing status, mode, rate, target volume, accumulated 

volume, target volume, sound volume, pressure, department, 

bed No., I/V set, temperature of medication

 

 

Blood Pressure and SPO2 Part

 

NIBP Range

Average deviation is not more than 5mmHg

Standard deviation is less than or equal to 8mmHg

Adult SYS 40mmHg~280mmHg

MEAN 20mmHg~240mmHg

DIA 10mmHg~220mmHg

 

Pediatric SYS 40mmHg~220mmHg

MEAN 20mmHg~170mmHg

DIA 10mmHg~160mmHg

 

Neonate SYS 40mmHg~135mmHg

MEAN 20mmHg~110mmHg

DIA 10mmHg~100mmHg

 

Static pressure Range 0mmHg~300mmHg

Error 3mmHg

 

Resolution: 1mmHg

 

NIBP measurement model

Manual

Cycle(automatic):according to the set time interval automatically start the NIBP measurement, the time interval of optional: 1min, 2min, 3min, 4min, 5min, 10min, 15min, 30min, 60min, 90min

 

Fast:continuous NIBP measurement, interval is 5min

NIBP alarm

High pressure alarm: >90mmHg-160mmHg

Low pressure alarm: <50mmHg-90mmHg

Accuracy:±5%

 

Overvoltage protection

Adult 280 mmHg±10 mmHg

Children 237 mmHg±3 mmHg

Newborn 147 mmHg±3 mmHg

 

SPO2

Range 0%~100%

Resolution 1%

Accuracy 70% to 100%, precision is 2%

0% to 69%, its precision is not defined

 

SPO2 alarm

Upper limit 100%

Lower limit 90%

Alarm numerical accuracy 3%

Alarm time accuracy 2s

 

Pulse rate

Range 20bpm~300bpm

Resolution 1bpm

Accuracy 3bpm

 

Pulse rate alarm

Upper limit 120bpm

Lower limit 50bpm (Alarm value is the default infusion pump)

Alarm numerical accuracy 5%

Alarm time accuracy 2s

 

NIBP Cuff requirement

Human body contact parts materials Waterproof cloth

Size 515mm(L)×145mm(W), ±5%

 

Finger clip type blood oxygen sensor

Type:Finger clip type

Wire length:3000mm, ±5%

Probe body contact parts materials:Silica gel

Back Up