Products

Patient Monitor

 

 

SUN-9000A is a high-end fetal monitor, and it offers the most advanced integrated monitoring for the delivery room in big hospital; private obstetrician's office; antepartum clinic. They are designed for use in moving situation covers the entire continuum of antepartum and intrapartum applications.

Features  

12.1" Color TFT screen folding up to 90 degree

Interface of parameters of FHR, TOCO, FM with digital and curve display

24 hours data storage which can be playback and printed
Built-in 152mm long life thermal printer, the life exceeds over 20 years
Advanced DSP technology, realtime identify and measurement for FHR, accurate and reliable results
Battery indication, pregnant No. Display and Special high sensitive water proof probe
Many alarm functions are included: The monitoring time over alarm, no paper alarm
Wide range of applying voltage (100-240 V)
Manual / automatic fetal movement detection
3 in 1 transducers, avoid twisting
Compatible with obstetric central monitoring system


Optionals  
Twins monitoring
Maternal SpO2, NIBP, PR, ECG, TEMP, RESP

Specification 
Transducer: Multi-crystals, pulsed doppler , high sensitive
Strength: <5mW/cm2
Working frequency: 1.0MHz
Signal processing: special DSP system and modern recognization.
Measurement range: 30~210 bpm
Alarm Range:
High limit: 160,170,180,190 bpm
Low limit: 90,100,110,120 bpm
Maximum audio output power : 1.5 Watt
Toco: Measurement range: 0~100 units
Resolution: 1
Fetal Movement: Button Hand Count ( Automatic cumulative number of movements
Maternal SpO2 measure: Measure scope : 70%~99%. Measure accuracy: ±3%
Pregnant HR Measure: Measure scope: 30bpm~240bpm. Measure accuracy: ±2 bpm
NIBP measure:
a) NIBP-SYS (6.7~32.0Kpa) (50~240mmHg)
b) NIBP-DIA (3.4~26.6Kpa) (25~200mmHg)
c) NIBP-MEAN (2.0~24.0ka (15~180mmHg)
NIBP Accuracy: ±1.1kPa (±8mmHg) or ±5% of the results, take the bigger one.
Maximum Mean error: ±5mmHg
Maximum Standard deviation 8mmHg
Pulse Rate Accuracy: Maximum Mean error: ±2bpm
Measure accuracy: ±2 bpm or ±5% of the results, take the bigger one.
Measure mode: Auto, Manual
Power: a.c.100-240V,50/60Hz or 8.4 Li-LON
Power Consumption: < 20W
Dimension: 295W×240D×73H(mm)
Net weight: 1.75kg
Working Environment: Temperature: 5~40°C. Relative humidity: 10~80%