Products

Patient Monitor

 Instruction                                                                                                                                                            

SUN-9000J is a high-end fetal monitor, and it offers the most advanced integrated
monitoring for the delivery room in big hospital; private obstetrician's office; antepartum
clinic. They are designed for use in moving situation covers the entire continuum of
antepartum and intrapartum applications.

 

Features   
7" Color TFT screen folding up to 90 degree
Interface of parameters of FHR, TOCO, FM with digital and curve display
12 hours data storage which can be playback and printed
Built-in 152mm long life thermal printer, the life exceeds over 20 years
Advanced DSP technology, realtime identify and measurement for FHR, accurate and reliable results
Battery indication, pregnant No. Display
Special high sensitive water proof probe
Multi-crystals, wide beam form high sensitivity ultrasound transducer, 1MHz
pulsed wave, low ultrasound power, safer to the fetus
Fetal acoustic stimulator function, which can easily and safely wake up the sleeping fetus when monitoring
A standard patient event marker and a clinical event marking button to separately mark clinical events
Many alarm functions are included: The monitoring time over alarm, no paper alarm
Wide range of applying voltage (100-240 V)
Manual / automatic fetal movement detection
3 in 1 transducers, avoid twisting
Compatible with obstetric central monitoring system

 

Optionals   
Twins monitoring
Maternal SpO2, NIBP, PR, ECG, TEMP, RESP

 

Specification   
Fetal Heart Rate:
Transducer: Multi-crystals, pulsed doppler , high sensitive
Strength: <5mW/cm2
Working frequency: 1.0MHz
Signal processing: special DSP system and modern recognization.
Measurement range: 30~210 bpm
Alarm Range:
High limit: 160,170,180,190 bpm
Low limit: 90,100,110,120 bpm
Maximum audio output power : 1.5 Watt
Toco: Measurement range: 0~100 units
Resolution: 1
Fetal Movement: Button Hand Count ( Automatic cumulative number of movements)
Maternal SpO2 measure: Measure scope : 70%~99%. Measure accuracy: ±3%
Pregnant HR Measure: Measure scope: 30bpm~240bpm. Measure accuracy: ±2 bpm
NIBP measure:
a) NIBP-SYS (6.7~32.0Kpa) (50~240mmHg)
b) NIBP-DIA (3.4~26.6Kpa) (25~200mmHg)
c) NIBP-MEAN (2.0~24.0ka (15~180mmHg)
NIBP Accuracy: ±1.1kPa (±8mmHg) or ±5% of the results, take the bigger one.
Maximum Mean error: ±5mmHg
Maximum Standard deviation 8mmHg
Pulse Rate Accuracy: Maximum Mean error: ±2bpm
Measure accuracy: ±2 bpm or ±5% of the results, take the bigger one.
Measure mode: Auto, Manual
Power: a.c.100-240V,50/60Hz or 8.4 Li-LON
Power Consumption: < 20W
Dimension: 295W×240D×73H(mm)
Net weight: 1.75kg
Working Environment: Temperature: 5~40°C. Relative humidity: 10~80%

1.jpg

 

9000J.jpg

Back Up