Products

Patient Monitor

 Introduction:                                                                                                                                                        

 This equipment can monitor such parameters as ECG,RESP,SpO2,NIBP,and TEMP. It integrates parameter measuring module,display and recorder in one device to form a compact and portable device. At the same time, its built-in replaceable battery provides convenience for patient moving. Modular designed, easy to maintain and replaced.

 

Features:   


15 inch high definition color TFT display, touch screen.
Standard 6-parameter : ECG, HR, NIBP, SPO2, RESP, TEMP.
13 kinds of Arrhythmia analysis.
2-4 hours battery power supply.
720 hours of trend data playback.
Sound, light, text alarm, volume adjustable.
Connected to central monitoring system .
Optional thermal recorder, ETCO2 module, IBP module,Tolley  .

Monitoring contents, scan speed, volume and output contents can be set optionally.

Function of NIBP review, storage for up to 2400 NIBP data.

Adopt digital SpO2 technology, which has strong anti-interference and antiweak filling capability.

Calculation of drug concentration.

Network: connecting with central station, other Bed observation and software updating. Connection mode: wireless and wired.

Built-in rechargeable battery for uninterrupted monitoring.

Print ECG, SpO2, RESP, BP and temperature data with one-key.

Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).

Analysis function for heart rate variability(HRV) (optional)

 

Performance:  

 

ECG

 

Lead Mode 3-lead and 5-lead are optional

Lead Selection I, II, III, avR, avL, avF, V

Wave 5-lead: 2 channels

3-lead: 1channel

Gain ×2.5mm/mV, ×5.0mm/mV, ×10mm/mV, ×20mm/mV

HR Measuring and Alarm Range

Range 15 ~ 300 bpm 

Accuracy ±1% or ±1bpm, which is greater

Alarm Accuracy ±2bpm

Resolution 1 bpm

CMRR

 

Monitor ≥ 100 dB

Surgery ≥ 100 dB

Diagnosis ≥ 60 dB

Bandwidth

 

Surgery 1 ~ 20 Hz(+0.4dB,-3dB)

Monitor 0.5 ~ 40 Hz(+0.4dB,-3dB)

Diagnosis 0.05~75Hz(+0.4dB,-3dB); 76Hz~150Hz(+0.4dB,-4.5dB)

Calibration Signal 1 mV (Vp-p), ±5%  Accuracy

ST Segment Monitoring

Measuring and Alarm Range -0.6 mV~ + 0.8 mV

ARR

ARR Detecting Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY,  R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

Alarm

Available

Review

Available

Scan Speed for ECG Waveform is adjustable

12.5mm/s accuracy ±10%

25mm/s accuracy±10%

50mm/s accuracy±10%

Respiration 

 

Method R-F(RA-LL) Impedance 

Differential Input Impedance >2.5 MΩ

Measuring Impedance Range 0.3~5.0Ω

Baseline Impedance Range 100Ω– 2500Ω

Bandwidth 0.3 ~ 2.5 Hz

Resp. Rate

 

Measuring and Alarm Range 0~120rpm

Resolution 1 rpm

Measuring Accuracy ±2 rpm

Alarm Accuracy ±3rpm

Apnea Alarm 10 ~ 40 S

NIBP

Method Oscillometry

Mode Manual, Auto, continuous 

Measuring Interval in AUTO Mode 1 /2 /3 /4 /5/ 10/ 15/ 30 /60 /90/120/240/480/960 Min

Measuring Period in Continuous Mode 5 Min

Measuring and Alarm Range 10~270mmHg

Alarm Type SYS, DIA, MEAN

Resolution

 

Pressure 1mmHg

Cuff Pressure ±3 mmHg

Accuracy ±10% or ±8mmHg, which is greater

Over-pressure Protection:

Adult Mode 315±10 mmHg

Pediatric Mode 265±10 mmHg

Neonatal Mode 155±10 mmHg

SpO2

Measuring Range 0 ~ 100 %

Alarm Range 0 ~ 100 %

Resolution 1 %

Accuracy   70% ~ 100% ±2%

0% ~ 69% unspecified

Pulse Rate(PR)

 

Measuring and Alarm Range 0~250bpm

Resolution 1bpm

Measuring Accuracy ±2bpm or ±2%, which is greater

Alarm Accuracy ±2bpm

TEMP

Channel dual-channel

Measuring and Alarm Range 0 ~ 50°C

Resolution 0.1°C

Accuracy ±0.1°C

Actualization Interval about 1 Sec.

Average Time Constant < 10 Sec.

Alarm responding Time ≤2min

EtCO2

Method Sidestream or Mainstream 
Measuring Range for CO2 0~150mmHg

Resolution for CO2:

0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
Accuracy for CO2:  0 – 40 mm Hg ±2 mm Hg
                    41 – 70 mm Hg ±5%
                    71 – 100 mm Hg ±8%
                    101 – 150 mm Hg ±10%
                    Respiration Rate>80BPM ±12%
                    AwRR Range 2~150 rpm
                    AwRR Accuracy  ±1BPM

                    Apnea Alarm Available

IBP

Channel dual-channel

Label ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measuring and Alarm Range -50~350 mm Hg
Resolution 1 mm Hg

Accuracy ±2% or 1mm Hg, which is greater

Dsplay Mode 12.1" color TFT LCD with high-resolution.

 

Power Supply 220V, 50Hz

 

Safety Classification class Ⅰ,type CF defibrillation-proof part

 

Physical Characteristic: Dimension 380×350×300(mm) Net weight 4.8Kg

 

Accessories: 

 

1) Adult SpO2 probe(5-pin) 2) Adult NIBP cuff 3) Extending tube for blood pressure 4) ECG lead

 5) ECG electrode 6) Temperature probe 7) Power cord 8) Thermal recording paper(optional)9) User Manual

 

 

SUN-M600K  SUN-M700K .jpg